Obviously, it's too late at that point to prevent the beam from hitting the patient, but you'll know that something went wrong and can lock the system until the problem is found.
Obviously, it's too late at that point to prevent the beam from hitting the patient, but you'll know that something went wrong and can lock the system until the problem is found.
[1] http://tommytoy.typepad.com/.a/6a0133f3a4072c970b0147e2ed7f8...
As an aside, TLDs and IVD systems are falling out of favor for patient dose monitoring because AAPM TG 62 (referenced in the link in your other post) is not directly applicable to IMRT and VMAT modalities, which are pretty common these days.
[1]http://www.sunnuclear.com/medphys/patientqa/epidose/epidose....
[2]http://www.sunnuclear.com/snc_site/solutions/patientqa/perfr...
p.s. If you are employed in Med Phys allow me to make a quick plug for my open source Med Phys QA database project: http://qatrackplus.com/