Careful with that as an assumption:
> In people younger than 55, Apple Watch’s positive predictive value is just 19.6 percent. That means in this group — which constitutes more than 90 percent of users of wearable devices like the Apple Watch — the app incorrectly diagnoses atrial fibrillation 79.4 percent of the time. (You can try the calculation yourself using this Bayesian calculator: enter 0.001 for prevalence, 0.98 for sensitivity, and 0.996 for specificity).
> The electrocardiogram app becomes more reliable in older individuals: The positive predictive value is 76 percent among users between the ages of 60 and 64, 91 percent among those aged 70 to 74, and 96 percent for those older than 85.
(Source: https://www.statnews.com/2019/01/08/apple-watch-iffy-atrial-...)
That being said, I will not at all dispute the ECG tools as very valuable.
https://www.ajmc.com/conferences/acc-2019/giant-study-sugges... - discusses the findings in the context of the Apple Health "Study" announced at one of their events.