I've worked in the LVAD and artificial heart industry, designed several.
You are right, that most modern LVADs are rotary pumps which are technically non pulsatile. I say technically because yes, if you test the pump on a bench of course it's not pulsatile. But once they're implanted, and working in concert with the native heart the net result is somewhere in between. You have more flow that you did before, less puksatile, but still pulsatile.
The implant guidance for most lvads calls for the clinicians to lower the speed setting of the lvad until they can see the aortic valve intermittently opening and closing on echo. If they set the setting too high and the valve stays open, or too low and it doesn't often open, it will calcify in one of those two positions.
If they follow this guidance, the net flow is pulsatile.
Onto your question.
Look up "pump head". This is what happens to the brain on truly non puksatile flow, like you get when your on extended periods of support on a heart lung bypass machine.
Check out a company "ventriflo" who is working on a pulsatile pump in this space to overcome these traditional challenges. I helped design this too.
Cheers!