You are incorrect. This circuit can potentially apply unsafe currents across the patient. A standard medical biopotential measurement circuit will have single-point failure protection to prevent conditions like this. This is required by standards like EC11, EC13, and 60601. Those systems are extremely safe.
Most hobbyist designs, like the one in this article, do not have such protection. Indeed, for this particular circuit, a simple static discharge, as might happen when you're applying the electrodes after walking over carpet, can fry the input amplifier, connecting it to the power rails. ESD (and, indeed, defibrillation -- which is about a thousand times more energy than ESD) protection would be present and required in any medical biopotential amplifier, even unrelated to the single-point protection described above.
If the ESD failure happens to short out to power rails, your current will be your power rails divided by your electrode impedance. This distribution is described in EC11/EC13 (although electrode technology has improved since the days when those tests were done, so impedance is even lower). Even with the ancient electrodes in EC11/EC13, you would get unsafe currents some of the time. With modern electrodes, you would get unsafe currents much of the time.
I've seen dozens of circuits for home EEG/ECG/EMG. Almost all of them, including this one, are a deathtrap. They will work okay most of the time. At some point, someone's going to kill themselves.
Disclaimer: I am not a physician, but I've designed ECG circuits all the way through safe, successful human trials.